Main functions and features
●This instrument adopts a six-oil-cup split structure (control box and oil cup box are separate). The microprocessor automatically completes operations such as voltage boosting, holding, stirring, static settling, calculation, and printing.
●It uses a liquid crystal display with Chinese menu prompts, making operation simple.
●This instrument has overvoltage, overcurrent, limit, and other protection functions to ensure the safety of operators.
●This instrument can be equipped with small-gap oil cups according to user requirements (program provided separately).
●This instrument has strong anti-electromagnetic interference capability and can be used both in laboratories and outdoor field sites.
●It contains a sine wave generator and does not directly use the grid power as the test source, avoiding its influence on test results.
●Complies with GB/T507-2002 and DL/T429.9-1991 standards.
Technical specifications
Power supply voltage: AC 220V±22V
Power supply frequency: 50Hz±1Hz
Voltage distortion rate: <3%
Power: 200W
Output voltage: 0~60kV 0~80kV (optional)
Voltage boosting speed: 2.0, 2.5, 3.0, 3.5kV/s (adjustable)
Operating conditions
Ambient temperature: 0℃~+40℃
Relative humidity: